Breast imaging – Mammography technique – Pitfalls in techniques of mammography – Breast tissue inadequately pulled onto the image receptor

  

If the breasts are large they may not be completely included in the field of view in the cassette.

Probable cause – breast tissue inadequately pulled onto the image receptor
  • Large breasts may not be completely covered in the standard four-view mammogram. Additional views are necessary to make sure that the entire breast has been imaged. The clinical presentation and examination findings will help in the selection of additional views, such as XCCM, XCCL, or axillary tail views.


Rectification
  • Breast tissue that is not included in the CC and MLO views can be incorporated into additional views.

Inadequate positioning





Adequate views to include the right breast upper outer quadrant mass




Right breast XCCL and axillary tail view






























































  