Home / Training / Manuals / Atlas of breast cancer early detection / Learning

Atlas of breast cancer early detection

Filter by language: English / Русский

Breast imaging – Mammography technique – Basic functioning of mammography unit

  

  1. The breast to be examined is placed on the breast support and held in position with the compression plate.
  2. A high-voltage supply of electrons from the cathode collides with the rotating molybdenum anode to produce a characteristic radiation spectrum optimal for breast imaging.
  3. The X-ray photons exit the tube through a port window of beryllium and additional molybdenum filters to modify the spectrum.
  4. A molybdenum filter is used to remove the undesirable high-energy photons (more than 20 keV) from the spectrum. Only radiation with photon energies less than 20 keV reaches the breast tissue.
  5. These X-rays are then directed through the compressed breast tissue by a collimator.
  6. The X-ray beam passing through the breast is attenuated depending on the density of the tissues it encounters in the breast.
  7. The attenuated beam passes through the anti-scatter grid to a detector located on the opposite side. The detector is either a photographic film plate, which captures the X-ray image on film, or a solid-state detector, which transmits electronic signals to a computer to form a digital image of the breast.
  8. The AEC sensor at the base of the image receptor cassette measures the radiation passing through the breast and cassette. When a sufficient number of X-ray photons have reached the sensor, exposure is automatically terminated by the sensor.







































































  


Click on the pictures to magnify and display the legends



Click on this icon to display a case study

25 avenue Tony Garnier CS 90627 69366, LYON CEDEX 07 France - Tel: +33 (0)4 72 73 84 85
© IARC 2024 - Terms of use  -  Privacy Policy.