The lightweight (weight, ~250 g) compact variety of thermal ablator is powered by a rechargeable lithium-ion battery. The machine has a built-in light to visualize the cervix and an activation button to switch it on and off. The machine can be pre-set to different treatment timings (20 s, 30 s, or 40 s) and the desired temperature (100 °C). The procedure timing is automated, which means that treatment automatically stops at the end of the pre-set time and there is a visual and audible indicator. The machine includes multiple reusable probes with tips of 16 mm and 19 mm in diameter. The tip may be flat or have a small projection (nipple) at its centre. The probe can be fitted to the machine by the probe connector, which is covered with a silicone cap when the machine is not in use. A battery charger is included in the pack.
Setting up the battery-operated ablator:
Step 1: Charging the battery
- Plug the charger into a power outlet.
- The green “power” LED illuminates when the power is turned on.
- Place the battery into the charging base. The red LED illuminates, indicating that the battery is being charged.
- The red LED will turn off when the battery is fully charged.
- A completely discharged battery takes about 2 hours to be fully recharged.
- Remove the battery from the charger, and disconnect the charger from the power outlet.
- The battery may be left to be charged overnight without any damage.
Step 2: Getting the machine ready
Insert a charged battery into the handle of the unit by pushing the battery into place until you feel the locking tabs snap (the battery can be inserted only when correctly orientated). Select the probe of the desired size, and remove the silicone cover from the connector. Insert the probe connector into the front of the unit. Ensure that the probe is tightly inserted. If the probe is not securely connected, the machine will power on briefly and shut itself off.
Step 3: Turning the machine on
Turn on the machine by pressing the ON/OFF button located on a panel on the handle. The green LED on the panel and the white illumination LED at the front are turned on. One blue light on the panel will flash, indicating that unit is ready to be used (the tip is not heated yet).
Step 4: Heating the probe
Press the ON/OFF button a second time to start heating the probe tip. Four blue timer LEDs on the panel will flash for a few seconds, indicating that the tip has begun heating. When the tip has heated to 100 °C (within a few seconds), the blue timer LEDs stop flashing and a single beep is audible. Treatment can now begin.
Step 5: Treatment cycle
Once the treatment cycle is running, the blue timer LEDs turn off with an audible beep, one at a time, after each quarter of the procedure has finished. When all four blue timer LEDs are turned off, a longer beep is audible, indicating that the unit is no longer applying heat and has commenced its cool-down cycle. Once the cool-down cycle is complete, the front white LED lights will turn off, and the probe may then be removed from the treatment area. If a second treatment area is indicated, repeat the above steps before removing the probe.